| Availability: | |
|---|---|
| Quantity: | |
Overheat protection: Automatically cutting off heating with audible and visual alarming when the autoclave is overheated.
Electric interlock safety: Preventing from starting a cycle if door is not properly locked or from door opening if the chamber is pressurized or the power is disconnected.
Door-temperature interlock safety: Preventing from explosion of liquids if the liquids temperature is not cooled down the safe limit (20℃ lower than the local boiling point).
Anti-burning protection: Automatically cutting off heating with audible and visual alarming when the autoclave is lack of water.
Steam pressure safety valve: Automatically works when the chamber pressure exceeds the limit.
Electrical circuit safety device for over-current and shortcut protection.
Automatic operation of sterilization (heating, sterilization, steam discharge and drying) can be easily set and monitored by the LCD touch screen.
Built-in steam generator prevents the media from contamination.
Buzzer for working errors and the end of the cycle.
Automatic atmospheric pressure adjustment for the installation and operation in plateau areas.
Optional printer, real-time record sterilization process, error and alarm information.
| Model | ST-H60 | ST-H80 | ST-H100 | ST-H135 | ST-H150 | ST-H185 | ST-H200 | ST-H300 | ST-H450 |
| Capacity | 60L | 80L | 100L | 135L | 150L | 185L | 200L | 300L | 450L |
External Size (WxDxH)(mm) | 985x600x1450 | 985x600x1450 | 1200x600x1450 | 1000x710x1550 | 1330x710x1550 | 1250x710x1550 | 1320x710x1550 | 1350x900x1750 | 1750x900x1750 |
Chamber Size (ØxH)(mm) | Ø386x500 | Ø386x700 | Ø386x950 | Ø500x700 | Ø500x780 | Ø500x950 | Ø500x1000 | Ø650x950 | Ø650x1360 |
Chamber Material | SUS304 | ||||||||
Sterilization Temperature | 105~136 ℃ | ||||||||
Designed Temperature | 150℃ | ||||||||
Max.Working Temperature | 136 ℃ | ||||||||
| Temp.Accuracy | 0.1℃ | ||||||||
| Temp.Uniformity | ≤±0.5℃ | ||||||||
Designed Pressure | 0.3Mpa | ||||||||
Sterilization Pressure | 0.23Mpa | ||||||||
| Temp. Controller | PID control by microprocessor | ||||||||
| Display | LCD touch screen | ||||||||
| Drying Function | YES | ||||||||
| Drying Time | 15min | ||||||||
Sterilization Program | ●Dressings/Fabric program ●Medical instrument program ●Rubber program ●Liquid program ●DIY program | ||||||||
Power Consumption | 5.7KW | 5.7KW | 6KW | 9.35KW | 9.35KW | 10.8KW | 10.8KW | 12KW | 20KW |
| Power Supply | 380V 50Hz | 380V 50Hz | 380V 50Hz | 380V 50Hz | 380V 50Hz | 380V 50Hz | 380V 50Hz | 380V 50Hz | 380V 50Hz |
Standard Sterilization Shelf | 1pcs | 1pcs | 1pcs | 1pcs | 1pcs | 1pcs | 1pcs | 1pcs | 1pcs |
Optional Accessories | Printer | ||||||||
| Net Weight | 105kg | 120kg | 140kg | 225kg | 240kg | 255kg | 265kg | 350kg | 450kg |
| Gross Weight | 165kg | 185kg | 210kg | 315kg | 340kg | 365kg | 380kg | 470kg | 570kg |
Shipping Size (WxDxH)(mm) | 1280x700x1600 | 1280x700x1600 | 1450x700x1600 | 1300x850x1900 | 1400x900x1900 | 1440x930x2020 | 1490x930x2020 | 1650x1100x2120 | 2100x1100x2120 |
Standard models start at 100L capacity and can process large quantities of dressings, culture media, and medical devices in a single cycle, making them ideal for bulk sterilization in hospitals, large laboratories, food processing plants, and similar settings.
Most models feature pulsed vacuum degassing, which enhances steam penetration. This ensures more thorough sterilization of porous and wrapped items, resulting in a pass rate significantly higher than that of standard vertical models.
Drain all condensate after each use, regularly clean the O-rings and the inner walls of the chamber, calibrate the pressure and temperature sensors monthly, and replace the safety valve on schedule to prevent safety hazards.