| Availability: | |
|---|---|
| Quantity: | |
UV Fluorescence Total Sulfur Analyzer complies strictly with ASTM D5453. Masterfully integrating high-sensitivity photomultiplier detection technology with an intelligent, modern control system, FSO-X1106 deliver rapid, environmentally friendly and highly accurate trace sulphur analysis solutions specifically for the petrochemical refining industry, third-party testing organisations and research institutes.
Integrated system comprising core components including main unit, temperature/gas flow control unit, reaction chamber, sample injector, computer and printer.
USB communication for plug-and-play functionality and stable data transmission.
High-specification, 24-bit high-precision data acquisition chip; capable of capturing weak signals with a minimum detection threshold of less than 1 μV.
Utilizes original imported components for key core parts.
Photomultiplier tube (PMT) high-voltage level is fully adjustable to meet specific detection requirements.
Highly flexible standard sample curve calibration, supporting both single-point and multi-point calibration methods.
Delivers exceptional detection sensitivity, superior linearity, and outstanding anti-interference capabilities.
Ensures accurate, stable, and reliable analysis results for even the most complex petroleum or chemical samples.
Powerful supporting software with excellent environmental adaptability, compatible with all mainstream operating systems currently on the market.
Comprehensive data file management, allowing for the storage, retrieval, searching, and printing of test data within the system at any time.
Standard compliance: ASTM D5453.
Model | FSO-UV4660 |
Technology | Ultraviolet Fluorescence (UVF) via High-Temperature Combustion |
Sample | Liquids, Solids, and Gases (Requires corresponding injector modules) |
Analytical Range | Liquid Samples: 0.2 mg/L to 10,000 mg/L; |
Solid Samples: 1.0 mg/L to 10,000 mg/L; | |
Gas Samples: 0.5 mg/m³ to 10,000 mg/m³; | |
(Note: Dilution is required for samples with concentrations > 10,000 mg/L) | |
Measurement Accuracy | For 0.2 mg/L ≤ X < 1.0 mg/L: Absolute Error ≤ ± 0.2 mg/L; |
For 1.0 mg/L ≤ X < 10 mg/L: RSD (Cv)≤ 10%; | |
For X > 10 mg/L: RSD (Cv)≤ 5%; | |
Gas Supply | High-Purity Oxygen (O₂): ≥ 99.98%; |
High-Purity Argon (Ar): ≥ 99.98%; | |
Furnace Temp. Range | Ambient to 1300°C |
Temp. Control Precision | ± 1°C |
Data Interface | USB |
Power Supply | AC 220±10%V, 50/60Hz |
Consumption | 2000 W |
External Size (W×D×H) | 650×550×420 mm |
Shipping Size (W×D×H) | 30×620×500 mm |
N.W./G.W. | 66/78 kg |
Operating Temp. and Humidity | 5–40 ℃, ≤85% RH |