Availability: | |
---|---|
Model | ELMW-812 |
---|---|
Plate models | Flat-, round -, U-, V- plate (96-well and 48-well) |
Manifold types | 8- and 12- way |
Wash times | 1 to 99 times adjustable |
Wash lines | 1~12 lines selectable |
Residual volume | <0.7 μl /well |
Wash mode | plates or strips |
Dispensing uniformity | CV < 1.5% |
Dispensing volume | 50~950 μl, interval 50 μl adjustable(May extend to 0~6000 μl, interval 25 μl adjustable) |
Number of dispensing channels | 3 channels |
Soaking and shaking time | 0~999 (seconds/minutes/hours) adjustable |
Aspiration time | 0.1~9.9 seconds adjustable, interval 0.1second |
Pipeline wash time | 0~240 seconds adjustable |
Distilled water automatically wash interval | 0-20 plates |
Operation language | English |
Power Supply | 220/110V,50/60Hz |
Shipping size(W×D×H)(mm) | 665x550x540 |
N.W./G.W. (kg) | 29/35 |
3.5-inch LCD, and suitable for flat, U and V-bottom microplates.
Easy to name, edit and save different 100 wash programs.
Any line and hole can be easily cleaned, unnecessary to complement the whole plate or strip. Detachable manifolds makes easy to clear particles in the manifolds. And special rinse procedure makes easy to maintain the device and reduces contamination between plates.
Stop or Pause key in the panel can be chosen to stop or pause during working.
Aspirating in 2 positions, residual volume <0.7μI.
Multi-plate can be selected by default settings or manual operation.
Microplate holder is detachable, which makes easy to disinfect and maintain it. And it's detachable, which makes easy to disinfect and maintain it.
Wash bottle filter keeps sediment in the solution away the pipes.
The control technology for pump pressure can avoid clogging in needles.
Automatical rinse procedure by distilled water when turn on or turn off can avoid crystallizing in the pipes.
Scientific structure design avoids bottom contaminating. Patent design of washer manifolds avoids it out of position or flop.
It is controlled by microcomputer to wash 48 well and 96 well microplate in experiments of enzyme linked immunosorbent assay (ELISA), Time-Resolved
Immunofluorometric assay (TRFIA) and chemiluminescence, which are widely used in clinical diagnostic laboratories, centers for disease control and prevention, animal and quarantine, animal husbandry and prevention, veterinary epidemic prevention stations, biotechnological industries, food industries, envi ronmental science, agricultural scientific research and other academic organizations.
Bottom holder is slope, in order that all waste water can be pumped out, which can prevent microplates from contaminating.
Program panel
Waste water sensor
Wash bottle filter
Model | ELMW-812 |
---|---|
Plate models | Flat-, round -, U-, V- plate (96-well and 48-well) |
Manifold types | 8- and 12- way |
Wash times | 1 to 99 times adjustable |
Wash lines | 1~12 lines selectable |
Residual volume | <0.7 μl /well |
Wash mode | plates or strips |
Dispensing uniformity | CV < 1.5% |
Dispensing volume | 50~950 μl, interval 50 μl adjustable(May extend to 0~6000 μl, interval 25 μl adjustable) |
Number of dispensing channels | 3 channels |
Soaking and shaking time | 0~999 (seconds/minutes/hours) adjustable |
Aspiration time | 0.1~9.9 seconds adjustable, interval 0.1second |
Pipeline wash time | 0~240 seconds adjustable |
Distilled water automatically wash interval | 0-20 plates |
Operation language | English |
Power Supply | 220/110V,50/60Hz |
Shipping size(W×D×H)(mm) | 665x550x540 |
N.W./G.W. (kg) | 29/35 |
3.5-inch LCD, and suitable for flat, U and V-bottom microplates.
Easy to name, edit and save different 100 wash programs.
Any line and hole can be easily cleaned, unnecessary to complement the whole plate or strip. Detachable manifolds makes easy to clear particles in the manifolds. And special rinse procedure makes easy to maintain the device and reduces contamination between plates.
Stop or Pause key in the panel can be chosen to stop or pause during working.
Aspirating in 2 positions, residual volume <0.7μI.
Multi-plate can be selected by default settings or manual operation.
Microplate holder is detachable, which makes easy to disinfect and maintain it. And it's detachable, which makes easy to disinfect and maintain it.
Wash bottle filter keeps sediment in the solution away the pipes.
The control technology for pump pressure can avoid clogging in needles.
Automatical rinse procedure by distilled water when turn on or turn off can avoid crystallizing in the pipes.
Scientific structure design avoids bottom contaminating. Patent design of washer manifolds avoids it out of position or flop.
It is controlled by microcomputer to wash 48 well and 96 well microplate in experiments of enzyme linked immunosorbent assay (ELISA), Time-Resolved
Immunofluorometric assay (TRFIA) and chemiluminescence, which are widely used in clinical diagnostic laboratories, centers for disease control and prevention, animal and quarantine, animal husbandry and prevention, veterinary epidemic prevention stations, biotechnological industries, food industries, envi ronmental science, agricultural scientific research and other academic organizations.
Bottom holder is slope, in order that all waste water can be pumped out, which can prevent microplates from contaminating.
Program panel
Waste water sensor
Wash bottle filter