Availability: | |
---|---|
Quantity: | |
Single Channel Pipette Function Display
A multifunctional motor-driven digital control pipette
All operational settings are controlled by two buttons.
Ergonomic design, low body bulk, and light weight make pipetting fatigue-free and easy to handle.
The only multifunction electronic pipette for dilution, mixing, stepping, and pipetting.
Using a basic electric pipette for pipetting and mixing
Variable aspiration and dispensing speeds
A Li-ion battery with two charging options allows for extended operation.
Self-adjusting
Channel | Volume Range | Increment | Test Volume | Error limits in accordance to ISO8655-2 | |||
---|---|---|---|---|---|---|---|
Accuracy error | Precision error | ||||||
μL | μL | μL | μL | % | s.d.*μL | CV*% | |
1 | 0.5-10 | 0.01 | 10 | ±0.10 | ±1.00 | 0.05 | 0.5 |
1 | ±0.035 | ±3.50 | 0.03 | 3 | |||
1 | 5-50 | 0.1 | 50 | ±0.40 | ±0.80 | 0.15 | 0.3 |
5 | ±0.15 | ±3.00 | 0.125 | 2.5 | |||
1 | 30-300 | 1 | 300 | ±1.80 | ±0.60 | 0.6 | 0.2 |
30 | ±0.90 | ±3.00 | 0.21 | 0.7 | |||
1 | 100-1000 | 5 | 1000 | ±6.00 | ±0.60 | 2 | 0.2 |
100 | ±3.00 | ±3.00 | 0.6 | 0.6 | |||
*s.d. = Standard Deviation *CV = Coefficient of Variation | Specifications are used as guidelines and the user calibration should refer to the industrial standard ISO 8655 |
Electronic control technology makes the pipetting process more accurate. The digital design ensures the precision of every drop of liquid to meet the high requirements of experiments.
The unique single-channel structure makes this electronic single channel pipette more convenient for single sample handling, ensuring efficient and individualised operation.
Offers a wide range of adjustable volumes, from microlitre to large volumes, suitable for a wide range of experimental needs.
Single channel pipette is equipped with a clear electronic display, intuitively presenting the current volume setting, so that users can easily grasp the working status.
Built-in rechargeable battery, support rapid charging, to ensure stable use for a long time.
Adopt ergonomic design, comfortable to hold.
The introduction of intelligent liquid handling technology, can achieve automatic liquid suction and discharge, reduce human error, improve the efficiency of the experiment.
Single Channel Pipette Function Display
A multifunctional motor-driven digital control pipette
All operational settings are controlled by two buttons.
Ergonomic design, low body bulk, and light weight make pipetting fatigue-free and easy to handle.
The only multifunction electronic pipette for dilution, mixing, stepping, and pipetting.
Using a basic electric pipette for pipetting and mixing
Variable aspiration and dispensing speeds
A Li-ion battery with two charging options allows for extended operation.
Self-adjusting
Channel | Volume Range | Increment | Test Volume | Error limits in accordance to ISO8655-2 | |||
---|---|---|---|---|---|---|---|
Accuracy error | Precision error | ||||||
μL | μL | μL | μL | % | s.d.*μL | CV*% | |
1 | 0.5-10 | 0.01 | 10 | ±0.10 | ±1.00 | 0.05 | 0.5 |
1 | ±0.035 | ±3.50 | 0.03 | 3 | |||
1 | 5-50 | 0.1 | 50 | ±0.40 | ±0.80 | 0.15 | 0.3 |
5 | ±0.15 | ±3.00 | 0.125 | 2.5 | |||
1 | 30-300 | 1 | 300 | ±1.80 | ±0.60 | 0.6 | 0.2 |
30 | ±0.90 | ±3.00 | 0.21 | 0.7 | |||
1 | 100-1000 | 5 | 1000 | ±6.00 | ±0.60 | 2 | 0.2 |
100 | ±3.00 | ±3.00 | 0.6 | 0.6 | |||
*s.d. = Standard Deviation *CV = Coefficient of Variation | Specifications are used as guidelines and the user calibration should refer to the industrial standard ISO 8655 |
Electronic control technology makes the pipetting process more accurate. The digital design ensures the precision of every drop of liquid to meet the high requirements of experiments.
The unique single-channel structure makes this electronic single channel pipette more convenient for single sample handling, ensuring efficient and individualised operation.
Offers a wide range of adjustable volumes, from microlitre to large volumes, suitable for a wide range of experimental needs.
Single channel pipette is equipped with a clear electronic display, intuitively presenting the current volume setting, so that users can easily grasp the working status.
Built-in rechargeable battery, support rapid charging, to ensure stable use for a long time.
Adopt ergonomic design, comfortable to hold.
The introduction of intelligent liquid handling technology, can achieve automatic liquid suction and discharge, reduce human error, improve the efficiency of the experiment.