Availability: | |
---|---|
Quantity: | |
Model | CON-P310 | |
---|---|---|
Conductivity | Range | 0.000 μS/cm to 1000 mS/cm |
Resolution | 0.001 μS/cm minimum, various with range selection | |
Accuracy | ±1.0% FS | |
Reference Temperature | 20, 25 ºC | |
Calibration Points | Up to 3 | |
Standard Recognition | 84μS/cm, 1413μS/cm, 12.88mS/cm | |
Resistivity | Range | 5.00 Ω·cm~20.00 MΩ ·cm |
Resolution | 0.01Ω·cm minimum, | |
Accuracy | ±1.0% FS | |
TDS | Range | 0.00ppm~300ppt |
Resolution | 0.001ppt minimum, 3 significant digits | |
Accuracy | ±1.0%FS | |
Salinity | Range | 0~80 ppt |
Resolution | 0.1 ppt | |
Accuracy | ±2 ppt | |
Temperature | Range | -5 to 110 ºC, 23 to 230 ºF |
Unit | ºC, ºF | |
Resolution | 0.1 | |
Relative Accuracy | ±0.2 | |
Measurement | Reading Mode | Auto Read (Fast, Medium, Slow), Timed, Continuous |
Reading Prompts | Reading, Stable, Locked | |
Temp. Compensation | ATC, MTC | |
Data Management | Data Storage | 1000 Groups |
GLP Features | Yes | |
Inputs | Conductivity with Temp. Probe | 5-pin aviation connector |
Outputs | USB | PC, printer |
Display Options | Backlight | Yes |
Auto Shutdown | 300, 600, 1200, 1800, 3600sec., off | |
IP Rating | IP65 | |
Date and Time | Yes | |
General | Electricity | Rechargeable Lithium battery, AC Adapter, 100-240V AC input, DC5V output |
Sizes(W×D×H) | 255×90×40 | |
Shipping Size (W×D×H)(mm) | 410×330×165 | |
N.W./G.W. (kg) | 0.5/5.5 |
CON-P310 is a standard portable conductivity meter that measures various solutions' conductivity accurately. The 4.3 inch LCD is the measurement data. The high accuracy of measurement can meet the needs of scientific research, environmental monitoring and other industries. It is the instrument of choice for pharmaceutical, water treatment and aquaculture industries.
Multiple reading function
Automatic/manual temperature compensation ensures accurate results.
Auto hold function senses and locks measurement endpoints.
Data capacity up to 500 groups (GLP compliant).
USB communication for easy data transfer
Auto Power Off function extends battery life.
Reset function; 1-3 point calibration
Waterproof and portable for water quality measurements in the field.
Can set parameters, including battery constant, temperature compensation coefficient, and TDS factor.
Model | CON-P310 | |
---|---|---|
Conductivity | Range | 0.000 μS/cm to 1000 mS/cm |
Resolution | 0.001 μS/cm minimum, various with range selection | |
Accuracy | ±1.0% FS | |
Reference Temperature | 20, 25 ºC | |
Calibration Points | Up to 3 | |
Standard Recognition | 84μS/cm, 1413μS/cm, 12.88mS/cm | |
Resistivity | Range | 5.00 Ω·cm~20.00 MΩ ·cm |
Resolution | 0.01Ω·cm minimum, | |
Accuracy | ±1.0% FS | |
TDS | Range | 0.00ppm~300ppt |
Resolution | 0.001ppt minimum, 3 significant digits | |
Accuracy | ±1.0%FS | |
Salinity | Range | 0~80 ppt |
Resolution | 0.1 ppt | |
Accuracy | ±2 ppt | |
Temperature | Range | -5 to 110 ºC, 23 to 230 ºF |
Unit | ºC, ºF | |
Resolution | 0.1 | |
Relative Accuracy | ±0.2 | |
Measurement | Reading Mode | Auto Read (Fast, Medium, Slow), Timed, Continuous |
Reading Prompts | Reading, Stable, Locked | |
Temp. Compensation | ATC, MTC | |
Data Management | Data Storage | 1000 Groups |
GLP Features | Yes | |
Inputs | Conductivity with Temp. Probe | 5-pin aviation connector |
Outputs | USB | PC, printer |
Display Options | Backlight | Yes |
Auto Shutdown | 300, 600, 1200, 1800, 3600sec., off | |
IP Rating | IP65 | |
Date and Time | Yes | |
General | Electricity | Rechargeable Lithium battery, AC Adapter, 100-240V AC input, DC5V output |
Sizes(W×D×H) | 255×90×40 | |
Shipping Size (W×D×H)(mm) | 410×330×165 | |
N.W./G.W. (kg) | 0.5/5.5 |
CON-P310 is a standard portable conductivity meter that measures various solutions' conductivity accurately. The 4.3 inch LCD is the measurement data. The high accuracy of measurement can meet the needs of scientific research, environmental monitoring and other industries. It is the instrument of choice for pharmaceutical, water treatment and aquaculture industries.
Multiple reading function
Automatic/manual temperature compensation ensures accurate results.
Auto hold function senses and locks measurement endpoints.
Data capacity up to 500 groups (GLP compliant).
USB communication for easy data transfer
Auto Power Off function extends battery life.
Reset function; 1-3 point calibration
Waterproof and portable for water quality measurements in the field.
Can set parameters, including battery constant, temperature compensation coefficient, and TDS factor.