It adopts special heating furnace to ensure the safety of test.
The heating power is continuously adjustable. It can meet requirements of test.
The open cup flash point tester is fully self-contained complete. The operator can do determination as long as connecting with coal gas or other civil gas.
With fast cooling function, it can improve test efficiency.
| Model | FPT-3536 |
|---|---|
| Heating Device | Furnace heating, no naked fire, explosion prevented. The power is adjustable from 0W to 600W |
| Test Flame Applicator | It applies the test flame automatically |
| Thermometer | -6~400℃ |
| Thermometer Scale | 2℃ |
| Igniting Device | Ignition source: coal gas (or civil gas) |
| Nozzle aperture: about 0.8mm | |
| Ambient Temperature | 15~35℃ |
| Relative Humidity | ≤85% |
| Power Supply | AC(220±10%)V, 50Hz |
| Power Consumption | ≤650W |
| Size(W×D×H)(mm) | 350×290×350 (thermometer is not included) |
| Net Weight | 5.5kg |
| Package Size (W×D×H)(mm) | 650x450x420 |
| Gross Weight | 19.6kg |
An open cup flash point tester is a laboratory instrument used to determine the flash point and fire point of petroleum products, lubricants, chemicals, and other flammable liquids. The flash point is the lowest temperature at which the vapor of a sample will ignite instantly upon contact with an open flame under specified test conditions. The fire point is the temperature at which the sample can sustain combustion for at least 5 seconds. This test is critical for safety classification, quality control, and regulatory compliance in industries such as petrochemicals, aviation, maritime transport, and power generation.