Availability: | |
---|---|
Quantity: | |
The low-temperature steam formaldehyde sterilizer uses a vacuum system to create a low-pressure environment in the sterilization room. In the low-pressure environment, the formaldehyde solution is vaporized to form low-temperature formaldehyde vapor. The formaldehyde vapor is used to act on the items in the sterilization room, causing microbial proteins and genetic substances on the items to denature and cause microbial death, in order to achieve the purpose of sterilization.
Mainly used for low-temperature sterilization of heat and moisture resistant items, it can be used in operating rooms, supply rooms, endoscopy rooms, and other treatment rooms in medical institutions.
Model | ST-FT100 | ST-FT130 | ST-FT150 |
---|---|---|---|
Capacity | 100L | 130L | 150L |
Chamber Size(WxDxH) | 700x430x360mm | 750x450x400mm | 820x645x400mm |
Consumption | 5.3KW | ||
Sterilizer Solution | 2% formaldehyde, 3% alcohol, and 95% distilled water solution | ||
Sterilizer Solution Packaging Specification | 2L | ||
Single Cycle Dosage | 0.6-1.3Kg | ||
Operating Temperature | 60℃ and 76℃ | ||
Sterilization Time | 2-3.5 hours | ||
Water Consumption | 3.95Kg/min | ||
Heating Method | Electric heating | ||
Anti-pressure | 5pa | ||
Operation Height | ≤160cm | ||
Chamber Material | 5052 corrosion resistant aluminum alloy, with antioxidant treated | ||
Cabinet Material | ABS | ||
Door Panel Material | 5052 corrosion resistant aluminum alloy, with antioxidant treated, ≥10mm | ||
Pipe Materials | SS304 stainless steel | ||
Insulation Materials | High quality insulation cotton | ||
Door Form | Automatic door | ||
Display | 8 inch Touch Screen | ||
External Size(WxDxH)(mm) | 1070x790x1770 | ||
Package Size(WxDxH)(mm) | 1200x900x1980 |
*Loading method: A movable stainless steel basket with support rods to avoid stacking items.
*Control system: Adopting Siemens PLC control system and friendly human-machine interaction interface, safe and reliable; One click start to finish, the entire process is automatically completed, displaying information such as temperature, pressure, sterilization time, etc. The system can be maintained and upgraded.
*Heating system: PID temperature controller is used to control the door, inner tank, and steam generator respectively and ensure their working stability.
*Display function: It can display temperature, pressure, time, operating mode, and alarm information.
*Power outage memory function: reduces operating costs and risk of misoperation.
*Steam generator: A self-developed fast evaporator that not only improves the vaporization rate of formaldehyde solution, but also enhances the sterilization effect.
*Cleaning and drying function: With the independent addition of automatic cleaning and drying function, it greatly reduces the formaldehyde residue in the processed items while ensuring the basic requirement of not getting wet, ensuring that it is below the national standard value.
*Automatic dosing function: Unique formaldehyde solution soft bag filling technology, automatic puncture filling, to avoid manual risks.
*Printing function: capable of printing program name, sterilization date, sterilization pot number, sterilization start and end time, as well as pressure, temperature, stage time and other information of each stage of the sterilization process. Printing records should be kept for more than 5 years.
*Storage function: Store sterilization data for no less than 15000 cycles, including program name, sterilization date, sterilization cycles, sterilization start and end time, as well as pressure, temperature, stage time, and end status of each stage of the sterilization process.
*Alarm system: There is a high temperature alarm system.
*Network connection: Equipped with a network connection port.
*Vacuum pump: Adopting a German water ring vacuum pump with reliable quality, it not only meets the design pressure requirements of the sterilizer, but also accelerates the decomposition and absorption of formaldehyde vapor to a certain extent, with a higher safety factor.
*Air filter: filtration accuracy ≤ 0.22 μ M;
*Pressure sensor: The built-in pressure sensor uses high-precision sensors to automatically detect the pressure inside the cavity, ensuring that the accuracy of the pressure inside the cavity is controlled within an effective range.
*Temperature sensor: embedded temperature sensor automatically detects the temperature inside the cavity, ensuring that the accuracy of the temperature inside the cavity is controlled within an effective range.
*Multiple protection functions: equipped with multiple protection functions such as overvoltage, undervoltage, overtemperature, door failure, and door safety interlock device
The low-temperature steam formaldehyde sterilizer uses a vacuum system to create a low-pressure environment in the sterilization room. In the low-pressure environment, the formaldehyde solution is vaporized to form low-temperature formaldehyde vapor. The formaldehyde vapor is used to act on the items in the sterilization room, causing microbial proteins and genetic substances on the items to denature and cause microbial death, in order to achieve the purpose of sterilization.
Mainly used for low-temperature sterilization of heat and moisture resistant items, it can be used in operating rooms, supply rooms, endoscopy rooms, and other treatment rooms in medical institutions.
Model | ST-FT100 | ST-FT130 | ST-FT150 |
---|---|---|---|
Capacity | 100L | 130L | 150L |
Chamber Size(WxDxH) | 700x430x360mm | 750x450x400mm | 820x645x400mm |
Consumption | 5.3KW | ||
Sterilizer Solution | 2% formaldehyde, 3% alcohol, and 95% distilled water solution | ||
Sterilizer Solution Packaging Specification | 2L | ||
Single Cycle Dosage | 0.6-1.3Kg | ||
Operating Temperature | 60℃ and 76℃ | ||
Sterilization Time | 2-3.5 hours | ||
Water Consumption | 3.95Kg/min | ||
Heating Method | Electric heating | ||
Anti-pressure | 5pa | ||
Operation Height | ≤160cm | ||
Chamber Material | 5052 corrosion resistant aluminum alloy, with antioxidant treated | ||
Cabinet Material | ABS | ||
Door Panel Material | 5052 corrosion resistant aluminum alloy, with antioxidant treated, ≥10mm | ||
Pipe Materials | SS304 stainless steel | ||
Insulation Materials | High quality insulation cotton | ||
Door Form | Automatic door | ||
Display | 8 inch Touch Screen | ||
External Size(WxDxH)(mm) | 1070x790x1770 | ||
Package Size(WxDxH)(mm) | 1200x900x1980 |
*Loading method: A movable stainless steel basket with support rods to avoid stacking items.
*Control system: Adopting Siemens PLC control system and friendly human-machine interaction interface, safe and reliable; One click start to finish, the entire process is automatically completed, displaying information such as temperature, pressure, sterilization time, etc. The system can be maintained and upgraded.
*Heating system: PID temperature controller is used to control the door, inner tank, and steam generator respectively and ensure their working stability.
*Display function: It can display temperature, pressure, time, operating mode, and alarm information.
*Power outage memory function: reduces operating costs and risk of misoperation.
*Steam generator: A self-developed fast evaporator that not only improves the vaporization rate of formaldehyde solution, but also enhances the sterilization effect.
*Cleaning and drying function: With the independent addition of automatic cleaning and drying function, it greatly reduces the formaldehyde residue in the processed items while ensuring the basic requirement of not getting wet, ensuring that it is below the national standard value.
*Automatic dosing function: Unique formaldehyde solution soft bag filling technology, automatic puncture filling, to avoid manual risks.
*Printing function: capable of printing program name, sterilization date, sterilization pot number, sterilization start and end time, as well as pressure, temperature, stage time and other information of each stage of the sterilization process. Printing records should be kept for more than 5 years.
*Storage function: Store sterilization data for no less than 15000 cycles, including program name, sterilization date, sterilization cycles, sterilization start and end time, as well as pressure, temperature, stage time, and end status of each stage of the sterilization process.
*Alarm system: There is a high temperature alarm system.
*Network connection: Equipped with a network connection port.
*Vacuum pump: Adopting a German water ring vacuum pump with reliable quality, it not only meets the design pressure requirements of the sterilizer, but also accelerates the decomposition and absorption of formaldehyde vapor to a certain extent, with a higher safety factor.
*Air filter: filtration accuracy ≤ 0.22 μ M;
*Pressure sensor: The built-in pressure sensor uses high-precision sensors to automatically detect the pressure inside the cavity, ensuring that the accuracy of the pressure inside the cavity is controlled within an effective range.
*Temperature sensor: embedded temperature sensor automatically detects the temperature inside the cavity, ensuring that the accuracy of the temperature inside the cavity is controlled within an effective range.
*Multiple protection functions: equipped with multiple protection functions such as overvoltage, undervoltage, overtemperature, door failure, and door safety interlock device