Availability: | |
---|---|
Model | TP-E300 |
---|---|
Technical Parameters | |
Electricity | AC220V 60Hz |
Main Fuse | 10A |
Power | 1000 VA |
Dry Weight, Unpacked | Approx. 160 Kg |
Weight | 210 Kg |
Temperature Range | 15 °C ~35 °C |
Relative Humidity | 10 % ~80 %, non-condensing |
Shipping Size(W×D×H)(mm) | 1290×880×710 |
N.W./G.W. (kg) | 180/230 |
Paraffin Wax Baths | |
Number of Baths | 3 |
Volume (L) | 4.0 l per bath |
Melt Time | Approx. 3 hrs |
Temperature | 65 °C |
Temperature Accuracy | ±1 °C |
Retort | |
Capacity | Approx. 300 cassettes |
Reagent Volume (In Liters) | 6.8 L |
Temperature (Paraffin) | 65 °C |
Temperature (Processing Reagents) | RT~35°C |
Temperature (Cleaning Reagents) | 65 °C |
Temperature Accuracy | ±1 °C |
Filling Time | 5 min |
Drain Time | 5 min |
Impregnation Vacuum | -70 kPa (g) |
Impregnation Pressure | 35 kPa (g) |
Fill Vacuum | -70 kPa (g) |
Drain Pressure | 35 kPa (g) |
General | |
Reagent Bottles | 10 |
Cleaning Solution Bottles | 2 |
Maximum Bottle Volume | 4.0 l |
Recirculation (Pump In/Out) | ON/OFF |
Time Between Cycles | 15 min |
This is Scitek's dedicated auto tissue processor for histopathology and routine clinics using the latest advanced technology having vacuum infiltration system. It features an innovative smart processor with woven paraffin infiltration, improving sample quality and laboratory economy. Proven enhancements and an intuitive user interface are the hallmarks of the TP-E300.
10 programs that consist of up to 10 reagent and 3 paraffin wax steps.
Time per program step: 0 to 99 hours, 59 minutes.
Up to 300 cassettes can be processed simultaneously.
Wax clean program.
10 reagent bottles.
3 wax baths.
2 cleaning solution bottles.
2 condensate bottle.
Vacuum infiltration system
Hardware and software
Large color touch screen LCD
User-friendly, intelligent software with password protection and provides report for each run
System check for the connection of Reagent and temperature sensors for both reagent and paraffin
Alarm system to reminder wrong operation and trouble/ errors
Multiple specimen protection system with spillage protections
Manual Operation during automation failure
Model | TP-E300 |
---|---|
Technical Parameters | |
Electricity | AC220V 60Hz |
Main Fuse | 10A |
Power | 1000 VA |
Dry Weight, Unpacked | Approx. 160 Kg |
Weight | 210 Kg |
Temperature Range | 15 °C ~35 °C |
Relative Humidity | 10 % ~80 %, non-condensing |
Shipping Size(W×D×H)(mm) | 1290×880×710 |
N.W./G.W. (kg) | 180/230 |
Paraffin Wax Baths | |
Number of Baths | 3 |
Volume (L) | 4.0 l per bath |
Melt Time | Approx. 3 hrs |
Temperature | 65 °C |
Temperature Accuracy | ±1 °C |
Retort | |
Capacity | Approx. 300 cassettes |
Reagent Volume (In Liters) | 6.8 L |
Temperature (Paraffin) | 65 °C |
Temperature (Processing Reagents) | RT~35°C |
Temperature (Cleaning Reagents) | 65 °C |
Temperature Accuracy | ±1 °C |
Filling Time | 5 min |
Drain Time | 5 min |
Impregnation Vacuum | -70 kPa (g) |
Impregnation Pressure | 35 kPa (g) |
Fill Vacuum | -70 kPa (g) |
Drain Pressure | 35 kPa (g) |
General | |
Reagent Bottles | 10 |
Cleaning Solution Bottles | 2 |
Maximum Bottle Volume | 4.0 l |
Recirculation (Pump In/Out) | ON/OFF |
Time Between Cycles | 15 min |
This is Scitek's dedicated auto tissue processor for histopathology and routine clinics using the latest advanced technology having vacuum infiltration system. It features an innovative smart processor with woven paraffin infiltration, improving sample quality and laboratory economy. Proven enhancements and an intuitive user interface are the hallmarks of the TP-E300.
10 programs that consist of up to 10 reagent and 3 paraffin wax steps.
Time per program step: 0 to 99 hours, 59 minutes.
Up to 300 cassettes can be processed simultaneously.
Wax clean program.
10 reagent bottles.
3 wax baths.
2 cleaning solution bottles.
2 condensate bottle.
Vacuum infiltration system
Hardware and software
Large color touch screen LCD
User-friendly, intelligent software with password protection and provides report for each run
System check for the connection of Reagent and temperature sensors for both reagent and paraffin
Alarm system to reminder wrong operation and trouble/ errors
Multiple specimen protection system with spillage protections
Manual Operation during automation failure