Availability: | |
---|---|
Quantity: | |
This blood gas and electrolyte analyser offers fully automated testing of ppH , pCO2, pO2, K+ , Na+, Cl- , Ca++, Hct in addition to blood gas parameters. The system offers individually adjustable parameter configurations for flexibility to meet your needs.
Model | BGE-95 | |
---|---|---|
Analyte | Measuring range | Unit |
Hydrogen Ion Activity (pH) | 6.000―9.000 | pH scale |
Carbon Dioxide | 0.667―26.67 | kPa |
Partial Pressure (pCO2) | 5.0―200.0 | mmHg |
Oxygen Partial Pressure (pCO2) | 0―106.7 | kPa |
0―800.0 | mmHg | |
Potassium Ion(K+) | 0.50―10.00 | mmol/L |
Sodiumion(Na+ ) | 50.0―200.0 | mmol/L |
Chloride Ion(Cl- ) | 20.0―200.0 | mmol/L |
Ionized Calcium (Ca2+) | 0.10―5.00 | mmol/L |
Hct | 12.0―65.0 | % |
B.P. | 500―800 | mmHg |
Glucose | 1.1―66.7 | mmol/L |
Lactate | 0.4―30.0 | mmol/L |
Derived(Calculated) Parameters: | ||
pH (TC) | pH temperature corrected | |
pCO2 (TC) | pCO2 temperature corrected | |
pO2 (TC) | pO2 temperature corrected | |
TCO2 | 0-50mmol/L | |
HCO3 | 0-50mmol/L | |
Beb | -25.0-25mmol/L | |
Beecf | -25.0-25mmol/L | |
SBC | 0-50mmol/L | |
SO2% | 40-100.0% | |
Ca2+(7.4) | Ionized calcium“normalized” to pH 7.4 | |
TCa | Total calcium content | |
AG | Anion gap | |
RI | 0.0-70.0 | |
AaDO2 | 0-700mmHg | |
O2SAT | Oxygen Saturation | |
tHb(c) | Hemoglobin (Calculated) | |
Sample | Whole blood, Serum, Plasma, Dialysate, CSF | |
Test Items | pH, pCO2, pO2, K+, Na+, Cl- , Ca++, Hct | |
Input Parameter | Patient temperature | |
Hemoglobin | ||
FiO | ||
B.P. | ||
Interface | Serial Line RS-232 | |
TCP/IP interface | ||
USB port | ||
Integrated Bar-code reader | ||
Analysis Time | <90s | |
Sample Volume | 95μL Whole blood | |
50uL Capillary blood | ||
Calibration | Automatic, programmable 1-point and 2-point calibration | |
Data Storage | >10000 | |
Display | 10.4” TFT touch screen | |
Battery | Nickel-metal hydride battery 3.0Ah | |
Printer | Build-in Thermal printer 70mm | |
Barcode Reader | Up to 14 kinds barcode type | |
Work Environment | 15℃-30℃ | |
Storage Environment | 5℃-40℃ | |
Power Supply | 100-240V, 50/60Hz | |
Cal Pak(WxHxD)(mm) | 258x180x70 | |
Autoqc(WxHxD)(mm) | 103x180x70 | |
Sizes(W×D×H)(mm) | 400x574x344 | |
Shipping Size (W×D×H)(mm) | 940x670x620 | |
N.W./G.W. (kg) | 30/60 |
The BGE-95 is an easy-to-use portable blood gas analyzer that provides accurate blood gas, electrolyte and critical care biochemistry results and is widely used in the healthcare industry.
Multiple combinations can test multiple parameters at once
Maintenance-free electrodes, saving maintenance cost and time
Capillary 35 UL, whole blood 95 UL, suitable for syringes and capillaries
Fast detection time can produce results in a short period
Automatic alarm in case of malfunction
Data storage >5000 items, you can easily view the monitoring data
High-resolution color touch screen, user-friendly operation interface
Built-in thermal printer for real-time printing of test data
The blood gas and electrolyte analyzer is equipped with a battery backup, even if the power outage can work normally
Scan the cassette bar code.
Remove the cassette from the package.
Place the cassette in the sample measurement chamber and close the lid.
Take a sample and view the data directly on the screen.
Large colour touch screen with clear display and friendly interface, users can easily carry out testing and data viewing.
This blood gas and electrolyte analyzer measures a number of key indicators such as pH, partial pressure of oxygen, partial pressure of carbon dioxide, sodium, potassium and calcium in a single test to meet clinical needs.
The electrolyte analyzer is compact and takes up little space, making it suitable for placement in a variety of laboratory and clinical environments.
Built-in data storage and export functions for easy transfer via USB and other interfaces for easy information management and record keeping.
A: Does the blood gas and electrolyte analyzer require regular maintenance?
Q: It does not require daily or weekly maintenance. However, regular calibration is required every three months.
A: Does the analyser have a QC lockout feature?
Q: Yes, the analyser ensures regulatory compliance by providing QC lockout.
This blood gas and electrolyte analyser offers fully automated testing of ppH , pCO2, pO2, K+ , Na+, Cl- , Ca++, Hct in addition to blood gas parameters. The system offers individually adjustable parameter configurations for flexibility to meet your needs.
Model | BGE-95 | |
---|---|---|
Analyte | Measuring range | Unit |
Hydrogen Ion Activity (pH) | 6.000―9.000 | pH scale |
Carbon Dioxide | 0.667―26.67 | kPa |
Partial Pressure (pCO2) | 5.0―200.0 | mmHg |
Oxygen Partial Pressure (pCO2) | 0―106.7 | kPa |
0―800.0 | mmHg | |
Potassium Ion(K+) | 0.50―10.00 | mmol/L |
Sodiumion(Na+ ) | 50.0―200.0 | mmol/L |
Chloride Ion(Cl- ) | 20.0―200.0 | mmol/L |
Ionized Calcium (Ca2+) | 0.10―5.00 | mmol/L |
Hct | 12.0―65.0 | % |
B.P. | 500―800 | mmHg |
Glucose | 1.1―66.7 | mmol/L |
Lactate | 0.4―30.0 | mmol/L |
Derived(Calculated) Parameters: | ||
pH (TC) | pH temperature corrected | |
pCO2 (TC) | pCO2 temperature corrected | |
pO2 (TC) | pO2 temperature corrected | |
TCO2 | 0-50mmol/L | |
HCO3 | 0-50mmol/L | |
Beb | -25.0-25mmol/L | |
Beecf | -25.0-25mmol/L | |
SBC | 0-50mmol/L | |
SO2% | 40-100.0% | |
Ca2+(7.4) | Ionized calcium“normalized” to pH 7.4 | |
TCa | Total calcium content | |
AG | Anion gap | |
RI | 0.0-70.0 | |
AaDO2 | 0-700mmHg | |
O2SAT | Oxygen Saturation | |
tHb(c) | Hemoglobin (Calculated) | |
Sample | Whole blood, Serum, Plasma, Dialysate, CSF | |
Test Items | pH, pCO2, pO2, K+, Na+, Cl- , Ca++, Hct | |
Input Parameter | Patient temperature | |
Hemoglobin | ||
FiO | ||
B.P. | ||
Interface | Serial Line RS-232 | |
TCP/IP interface | ||
USB port | ||
Integrated Bar-code reader | ||
Analysis Time | <90s | |
Sample Volume | 95μL Whole blood | |
50uL Capillary blood | ||
Calibration | Automatic, programmable 1-point and 2-point calibration | |
Data Storage | >10000 | |
Display | 10.4” TFT touch screen | |
Battery | Nickel-metal hydride battery 3.0Ah | |
Printer | Build-in Thermal printer 70mm | |
Barcode Reader | Up to 14 kinds barcode type | |
Work Environment | 15℃-30℃ | |
Storage Environment | 5℃-40℃ | |
Power Supply | 100-240V, 50/60Hz | |
Cal Pak(WxHxD)(mm) | 258x180x70 | |
Autoqc(WxHxD)(mm) | 103x180x70 | |
Sizes(W×D×H)(mm) | 400x574x344 | |
Shipping Size (W×D×H)(mm) | 940x670x620 | |
N.W./G.W. (kg) | 30/60 |
The BGE-95 is an easy-to-use portable blood gas analyzer that provides accurate blood gas, electrolyte and critical care biochemistry results and is widely used in the healthcare industry.
Multiple combinations can test multiple parameters at once
Maintenance-free electrodes, saving maintenance cost and time
Capillary 35 UL, whole blood 95 UL, suitable for syringes and capillaries
Fast detection time can produce results in a short period
Automatic alarm in case of malfunction
Data storage >5000 items, you can easily view the monitoring data
High-resolution color touch screen, user-friendly operation interface
Built-in thermal printer for real-time printing of test data
The blood gas and electrolyte analyzer is equipped with a battery backup, even if the power outage can work normally
Scan the cassette bar code.
Remove the cassette from the package.
Place the cassette in the sample measurement chamber and close the lid.
Take a sample and view the data directly on the screen.
Large colour touch screen with clear display and friendly interface, users can easily carry out testing and data viewing.
This blood gas and electrolyte analyzer measures a number of key indicators such as pH, partial pressure of oxygen, partial pressure of carbon dioxide, sodium, potassium and calcium in a single test to meet clinical needs.
The electrolyte analyzer is compact and takes up little space, making it suitable for placement in a variety of laboratory and clinical environments.
Built-in data storage and export functions for easy transfer via USB and other interfaces for easy information management and record keeping.
A: Does the blood gas and electrolyte analyzer require regular maintenance?
Q: It does not require daily or weekly maintenance. However, regular calibration is required every three months.
A: Does the analyser have a QC lockout feature?
Q: Yes, the analyser ensures regulatory compliance by providing QC lockout.